- Precise Viscosity Water Bath
• Maintaining uniform temperature inside the bath with inner circulation function
• Using high precision temperature sensor. Excellent temperature accuracy
• 10 steps flow rate control
• Precision control Implementation by 0.01°C
• Installing a drain tap for an easier drain from bath
• To control the temperature lower than the ambient temperature or to maintain the temperature precisely need to connect a separate cooler to a cooling line (optional)
• Lower temperature ranges are different depending on the capacity of the optional cooler connected
• Available to observe through tempered glass window installed front and rear of equipment
• Installed ports for external sensor and RS485 for PC communication
(communication program can be downloaded on www.lklab.com)
• Simple and intuitive user interface with graphic LCD display
• Auto tuning and temperature calibration function installed
• Automatic fault sensing system; bath itself can detect an unusual situation (patent registration)
• Low water alarm for heater's overheating avoiding due to lack of liquid inside a bath
• Double overheat safety device ‒ 1st controller overheat alarm, 2nd overheat shut off a circuit
• LK Labkorea
Cat. No | Model | Temp. Range | Accuracy |
---|---|---|---|
B04-01-220 | HVB40P | Ambient +10 ~ +100℃ | ±0.1℃ |
B04-01-225 | HVB40H | Ambient +10 ~ +150℃ |
- Optional Accessory
Cat. No | Model | Description |
---|---|---|
B04-01-085 | LI-USB | RS485 to USB Communication Cable |
B04-91-483 | ET101M | External Sensor, Pt100, Ø4×L300 mm |
Cat. No. | B04-01-220 | B04-01-225 | ||
---|---|---|---|---|
Model | HVB40P | HVB40H | ||
Capacity | 21 L | |||
Controller | Control | PID Control, Auto-tuning | ||
Display | GLCD (Graphic LCD) | |||
Resolution | 0.01℃ | |||
Timer | Run start / Wait start (99 hr 59 min) | |||
Temperature | Range | Ambient +10 ~ +100℃ | Ambient +10 ~ +150℃ | |
Accuracy | ±0.1℃ | |||
Uniformity | ±0.1℃ at 50℃ | |||
Size | Interior | w370×d200×h310 mm | ||
Exterior | w440×d275×h600 mm | |||
Top Open | w95×d95 mm, 2ea | |||
Electric Supply |
Power | 1 Phase, 220VAC, 50/60 Hz | ||
Max Consumption | 2.2 kW (10 A) | 2.5 kW (11.4 A) | ||
Power Line | Standard Plug | |||
Other | Drain | 1/4” Hose Nipple male | ||
Cooling IN / OUT | for 9 mm Silicone Tubing | |||
Safety | Over Temp. & Current Protection, Automatical Fault Sensing System |